WTW TriOxmatic 690 Online Dissolved Oxygen Sensor: Calibration & Membrane Care

Author: Zhejiang Weilan Technology Co., Ltd.Source: Official sitePublished: Category: Maintenance
Summary: Field guide for WTW TriOxmatic 690 online DO electrodes: water-saturated air calibration, electrolyte refilling, membrane cap replacement, and aeration basin mounting.

In municipal wastewater aeration basins, dissolved oxygen readings drive blower speed and energy use. The German WTW TriOxmatic 690 is a three-electrode polarographic online DO sensor known for low drift. Based on our WTW TriOxmatic 690 page, here is how to calibrate and maintain it.

1. Water-Saturated Air Calibration

  1. Rinse sludge and biofilm off the sensor, then gently blot water droplets from the membrane without rubbing.
  2. Hang the sensor inside a damp calibration sleeve or shaded humid air for 10 to 15 minutes until the temperature reading settles.
  3. Start CAL on the transmitter, which compensates for temperature and barometric pressure at 100% air saturation.

2. Membrane Cap and Electrolyte Care

Fill the new membrane cap with fresh OEM electrolyte, tap gently to release trapped air bubbles, and thread it onto the sensor body. Mount the probe at a 30 to 45 degree angle in the aeration basin so rising bubbles slide past, and polish tarnished gold cathodes only with the OEM polishing strips.

Conclusion

Bubble-free electrolyte fills and regular air-saturation checks keep the WTW TriOxmatic 690 accurate across changing basin loads. Contact our team for sensors, membrane caps, and electrolyte kits.

FAQ

Why must water droplets be blotted before air calibration?

Water droplets slow oxygen diffusion across the membrane and skew calibration equilibrium.

How often should the membrane cap be replaced?

Usually every 6 to 12 months in municipal wastewater, along with weekly rinsing to clear biofilm.

Why do DO readings drop continuously in stagnant water?

Polarographic sensors consume trace oxygen at the membrane surface, so the water needs enough flow velocity across the tip.

What causes persistently low or near-zero DO readings?

Check for a torn membrane, dried or bubbled electrolyte, or moisture inside the cable connector.