Mettler-Toledo InLab pH/ORP Electrodes: Selection, Activation & Cleaning
Mettler-Toledo InLab pH and ORP electrodes are widely used for lab titrations and field verification against online process transmitters. Based on our InLab electrode page, this guide covers junction choice, rehydration, and cleaning.
1. Choosing the Right Junction
Ceramic junctions work well in clean aqueous samples and standard buffers. If you measure wastewater, emulsions, or samples with suspended solids, pick a sleeve or open junction so particles do not clog the reference path.
2. Rehydrating a Dry Electrode
Do not calibrate a dried-out pH electrode right away. Soak the glass bulb and junction in 3 mol/L KCl (or pH 4.01 buffer) for 12 to 24 hours first. Never store pH electrodes in distilled or DI water, because pure water leaches potassium chloride out of the reference chamber.
3. Safe Cleaning by Foulant
- Inorganic scale or salts: soak 10 to 15 minutes in 0.1 mol/L HCl, then rinse with DI water.
- Oils or organics: rinse briefly in mild detergent or ethanol, then recondition in KCl solution.
- Never scrub the sensing glass bulb with abrasives or dry paper.
Conclusion
Storing electrodes in 3 mol/L KCl and cleaning with mild solutions keeps InLab pH/ORP slopes above 90%. Contact our team for replacement sensors or online analyzers.
FAQ
What solution should pH electrodes be stored in?
Always use 3 mol/L KCl storage solution, never pure or deionized water.
What is a healthy calibration slope?
Between 90% and 105% (about 54 to 62 mV/pH at 25 °C). A slope below 85% after cleaning usually means the glass membrane has aged out.
Can InLab electrodes connect to third-party transmitters?
Yes, if the connector (such as BNC) matches, though you should verify built-in temperature sensor compatibility (NTC vs PT1000).
Should ORP electrodes also be stored in KCl?
Yes. Combination ORP electrodes also depend on a wet reference junction and should stay moist in 3 mol/L KCl.
Do you supply both lab electrodes and industrial online meters?
Yes. Weilan supplies Mettler-Toledo InLab sensors alongside Suntex and ProMinent online analyzers.