Suntex PC-3310 Intelligent pH/ORP Transmitter (144×144 mm): Wiring & Calibration
For main control panels and water stations where operators read instruments from a distance, the Suntex PC-3310 offers a large 144×144 mm housing with IP65 front protection. Based on our Suntex PC-3310 and PC-3110 pages, here is how to wire and calibrate it.
1. Key Specifications
Suntex PC-3310 measures -2.00 to 16.00 pH and -1999 to +1999 mV (ORP) with automatic temperature compensation (PT1000/NTC), isolated 4–20 mA analog output, and programmable Hi/Lo/Wash relay contacts. Functionally it mirrors the 96×96 mm PC-3110 in a larger 144×144 mm format.
2. Terminal Wiring and Two-Point Calibration
Route high-impedance IN/REF electrode cables away from AC/VFD lines, connect PE earth ground securely to prevent drift, and wire the 4–20 mA output to your PLC or dosing pump. Calibrate with fresh pH 7.00 and pH 4.01 (or 10.01) buffers, keeping electrode slope between 85% and 105%.
Conclusion
Suntex PC-3310 combines high-visibility panel readout with reliable 4–20 mA and relay dosing control. Contact our team for panel cutout drawings or compatible sensors.
FAQ
What is the panel cutout for Suntex PC-3310?
The 144×144 mm bezel typically uses a 138×138 mm cutout; verify against the manual template before cutting.
How does PC-3310 differ from PC-3110?
They share the same measurement range, IP65 protection, and outputs; PC-3310 simply provides a larger 144×144 mm display.
Why is the reading stuck near 7.00 or pegged at full scale?
Check for moisture in the connector or a shorted coaxial cable, and run a zero-input short test on the transmitter.
Can PC-3310 trigger an automatic sensor wash valve?
Yes, the programmable wash relay drives air-blast or water-jet cleaning solenoids on a timer.
Do you supply Suntex transmitters and electrodes?
Yes. Weilan supplies PC-3310, PC-3110, PC-350, and PC-110 transmitters plus pH/ORP electrodes.