Zhejiang Weilan Technology Co., Ltd.

Online pH and ORP Controller Two-Point Calibration and Probe Cleaning Guide

Author: Zhejiang Weilan Technology Co., Ltd.Source: Official sitePublished: Category: News
Summary: Practical field guide for two-point calibration of industrial pH/ORP controllers (pH 6.86 + 4.00/9.18 buffers), glass bulb cleaning, and slope checks.

In automated acid-base neutralization and chlorine dosing skids, metering pumps follow whatever signal the inline analyzer sends. A fouled sensor or drifted pH controller calibration can shift readings by over 1.0 pH, causing severe chemical overdosing or discharge violations.

Follow this standardized two-point pH controller calibration and sensor cleaning routine used across industrial wastewater plants.

1. Buffer Selection and Pre-Calibration Cleaning

Use fresh pH 6.86 (zero offset) paired with pH 4.00 (acidic slope) for acid wastewater (pH 2–7), or pH 6.86 paired with pH 9.18 (alkaline slope) for caustic systems (pH 7–12).

Before touching buffers, pull the probe and rinse with deionized water. Remove carbonate scale by soaking 5 minutes in 3%–5% dilute hydrochloric acid, or wash grease off with mild detergent. Never scrub the glass sensing bulb with abrasives or rough paper.

2. Four-Step Two-Point Calibration Routine

  1. Enter CAL mode: Placing the transmitter into calibration mode normally freezes 4-20mA outputs (Hold) so connected metering pumps do not surge during testing.
  2. First point (pH 6.86): Swirl the rinsed probe gently in pH 6.86 buffer, wait 30–60 seconds for temperature-compensated stabilization, and confirm.
  3. Second point (4.00 or 9.18): Rinse thoroughly with deionized water, blot excess drops gently without wiping, and immerse in the second buffer until stable before confirming.
  4. Check sensor slope: Healthy glass electrodes show a slope between 95% and 102% (near 59.16 mV/pH @ 25 °C).

3. Signs the pH Electrode Needs Replacement

If post-calibration slope falls below 80%–85%, zero offset exceeds ±45 mV, or readings drift endlessly beyond 3 minutes in fresh buffer, the internal reference junction is depleted and the probe must be replaced.

Conclusion

Consistent probe cleaning and disciplined two-point pH controller calibration keep closed-loop dosing accurate and prevent wasted acid or caustic. Visit Products or Contact Us for industrial analyzers and replacement probes.

FAQ

Do ORP controllers also need two-point 4.00 and 6.86 calibration?

No. Industrial ORP probes are normally verified or single-point offset adjusted using a standard redox solution such as +256 mV or +220 mV.

Can used buffer poured into a cup be poured back into the stock bottle?

Never. Pouring used buffer back contaminates the entire bottle and ruins future calibrations.

Will dosing pumps turn on unexpectedly while calibrating?

Most industrial transmitters hold outputs steady during CAL mode, though switching dosing pumps to manual OFF beforehand is safest.

Why did a probe stop responding after sitting dry on a shelf?

The glass bulb requires a hydrated gel layer. Always store spare pH probes with 3.3 M KCl solution inside the protective cap, never dry or in pure distilled water.

Do you supply SEKO and SUNTEX inline pH/ORP controllers?

Yes, we supply complete controllers, chemical-resistant probes, flow cells, and buffer standards.